Step 17. Verify that the New Model Number Is Shown Correctly
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse is recommended for this procedure but is not required.
Mouse Procedure
1. On the analyzer's Help menu, click About Network Analyzer.
2. Verify that the new model number, E8363B, is shown correctly in the display. Click OK.
3. On the analyzer front panel, press Preset. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 40 GHz.
Front Panel Keys Procedure
1. In the COMMAND block, press Menu/Dialog.
2. In the NAVIGATION block, press the right tab and arrow keys to move over to the Help
menu, and down to the About Network Analyzer selection. Press Click.
3. Verify that the new model number, E8363B, is shown correctly, then press Click.
4. On the analyzer front panel, press Preset. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 40 GHz.
Installation Note E8362-90004