Step 9. Install the Remaining New Hardware
Install the A23 and A24 Detectors (Refer to
1. Place the A23 and A24 detectors on top of the brackets in the analyzer. Be sure to orient
them as shown.
2. Install two screws (item
3. Connect the wrapped wire control cables from the A23 and A24 detectors:
• A23 detector to A16 test set motherboard J5 (R1 DET)
• A24 detector to A16 test set motherboard J6 (R2 DET)
Install the A21 SOMA 50 (Refer to
1. Place the A21 SOMA 50 in position in the analyzer. Be sure to orient it as shown.
2. Install four screws (item
3. Connect the A21 SOMA 50 ribbon cable to A16 test set motherboard J4 (SOMA50).
Figure 13 A23 and A24 Detectors and A21 SOMA 50 Installation
20
Figure
x
) in each detector to secure them to the brackets.
Figure
13.)
y
) to secure the SOMA 50 to the A16 test set motherboard.
13.)
Installation Note E8362-90004