5.6 Attaching the touch screen display to the
column (new model)
1.
Unfasten the touch screen display arm (1 screw) from
vertical position and re-fasten it to the horizontal posi-
tion. One screw is pre-installed.
2.
Fasten the touch screen display to the arm by 2 screws.
3.
Connect the cables (Ethernet, PC power, the grounding
cable). The grounding point is indicated in the figure be-
low. Secure the cable with cable clamp on the touch
screen display bracket.