5 Installing the unit
Side horizontal balance in terms of the cephalostat
sensor.
54
11. Remove the upper cover (A) from the cephalostat head
and put it aside. Take the cephalostat head out of the
styrofoam packaging and lift it carefully to the hook on
the other end of the cephalostat arm.
A
12. The pin (A) goes through the inlet.
Attach the two bolts (B) to lock the cephalostat head to
its place.
13. Check the levelling of the cephalostat head from the
two spirit levels. If the cephalostat head is tilted or
slightly hanging, straighten it as instructed below.
Front horizontal balance in terms of the cephalostat sensor.
A
B
B