KaVo ORTHOPANTOMOGRAPH OP 3D Pro Installation Manual page 137

3d dental x-ray system
Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D Pro:
Table of Contents

Advertisement

Jumper
2.
Remove the cover (A) under the lower shelf.
3.
Remove the metal plate by loosening 4 screws.
4.
On the EA200 board there is a jumper (J1) for line volt-
age selection. If you have 220-240Vac mains voltage,
you should install jumper marked with 230Vac to the J1
position. If you have 100-110Vac mains voltage, you
should use the 100Vac jumper and if you have 110-
120Vac mains voltage, use the 120Vac jumper.
5.
Make sure that you have correct main fuses (F1 & F2)
in the unit.
8 Special procedures