(206908) is required (ordered separately). At least one
wall mount bracket is needed for all installations.
8.
Erect the column by lifting from the top. Move the unit
to the place where it is installed and set it beside the
wall.
HEAVY OBJECT 87 KG 191.8 LBS
A minimum of two persons are required for
the following task.
NOTICE! Follow the local regulations considering lifting
heavy objects.
CAUTION! Do not lift the unit from the patient handles!
CAUTION! Always support the unit before it is attached
to the wall! The unit will not stay in the upright position
unsupported!
9.
Remove the Styrofoam supports.
10. Attach the unit to the upper wall bracket on the wall with
screws and washers.