8 Special procedures
138
sions.txt included in the software release package. If all
versions match, proceed to the next step. Otherwise,
you must first update the mismatching components.
2.
Set the software release version by pressing F9, or by
manually typing:
pmode 0 <ver_major> <ver_minor>
Example:
pmode 0 2 00
The unit should reply
programming ended successfully, version now
R2.00
This command makes the unit store the versions of the
software components currently loaded, so that the va-
lidity of the software release may be verified later. If one
or more software component are manually changed
without this step being carried out, the version will be
displayed with an "M" postfix (as in "Modified"), e.g.
R1.11M.

8.8 Volatile installation

Volatile installation is useful when a circuit board is not
responsive to Ethernet communication, for example due to
corrupted software. Using a JTAG programmer the
software may then be loaded directly into the RAM of the
circuit board, after which the software may be permanently
installed the normal way through Ethernet. Volatile
installation is available for all embedded software
components except the GUI.
To perform a volatile installation you will need:
An Altera USB-Blaster Download Cable, available from
Altera at:
http://www.altera.com/products/devkits/kit-cables.html
Alternatively, a Terasic USB Blaster Download Cable,
available from Terasic can be used:
www.terasic.com.tw
Quartus II Programmer software, available for download
from Altera at: