EL6010-Caldos17: Measuring Range

Measuring Range
Step Interval
Reference
Measuring Range
Measuring Range
Notes
Active Sample
Component
42/24-30 EN Rev. 1
Measuring range
0.0 -
100.0 Vol%
Measuring range
0.0
-
100.0 Vol%
Measuring range
0.0 -
100.0
Vol%
The measuring range adjustment step interval is 0.5 Vol.-%.
Calibration of the measuring range should be verified after changing measuring
range limits.
The gas analyzer has a physical measuring range ("reference measuring range")
for each sample component. The limits of this range are 0 Vol.-% and 100 Vol.-%.
The measuring range can be adjusted within the limits valid for the sample
components (see Section "EL6010-Caldos17: Metrological Data", page 57). For
the factory-set component "relative thermal conductivity" it is rTC = 0 to 60,000.
The largest measuring range is 0 to 100 Vol.-% or 0 Vol.-% to saturation,
depending on the measuring task.
Measuring ranges within ignition limits cannot be provided.
Active TC component
CO2:H2
Active TC component
CO2:H2
All sample components are precalibrated at the factory. After activating a sample
component, check the zero and span points and recalibrate as necessary (see
Chapter "Calibrate the Gas Analyzer", page 35).
When a different sample component is activated, the gas analyzer reverts to the
factory precalibrated state. This means that calibration values are lost and set
point and measuring range limit settings are reset to the reference measuring
range limits.
EL6010 Gas Analyzer in Category 2G Operator's Manual
Measuring range
Measuring range start
Value range: Within the reference measuring range
Measuring range start
Value range: Within the reference measuring range
Active sample component
All the sample components listed in the Section
"EL6010-Caldos17: Metrological Data" are available
(see page 57).
31