Preparation For Installation; Installation Location Requirements; Prepare The Installation - ABB EL6010 Operator's Manual

Continuous gas analyzers
Table of Contents

Advertisement

Chapter 1

Installation Location Requirements

Installation Location
Short Gas Paths
Adequate
Air Circulation
Protection from
Adverse Conditions
Environmental
Conditions
8

Preparation for Installation

The gas analyzer should not be mounted outdoors.
Install the gas analyzer as close as possible to the sampling location.
Locate the gas conditioning and calibration assemblies as close as possible to the
gas analyzer.
Provide for adequate natural air circulation around the gas analyzer. Avoid heat
buildup.
Protect the gas analyzer from:
• Cold
• Direct sunlight and heat
• Large temperature variations
• Strong air currents
• Accumulations of dust and dust infiltration
• Corrosive atmospheres
• Vibration
Air pressure range
Relative humidity
Ambient temperature
Note: The explosion protection is not impaired if the gas analyzer is operated at
temperatures less than +5 °C and down to –20 °C. However in this temperature
range the compliance with the metrological data cannot be guaranteed.
EL6010 Gas Analyzer in Category 2G Operator's Manual
during storage and shipping
in operation
EL6010-Caldos17
EL6010-Magnos106
EL6010-Uras14
600 to 1250 hPa
max. 75 %
–25 to +65 °C
+5 to +50 °C
+5 to +50 °C
+5 to +45 °C
42/24-30 EN Rev. 1