Troubleshooting - ABB EL6010 Operator's Manual

Continuous gas analyzers
Table of Contents

Advertisement

Troubleshooting

Flow Fault
Gas lines or filter
dirty, plugged or
leaking
Analyzer unit gas
paths crimped or
leaking
Outlet pressure not
equal to atmospheric
pressure
Unstable Measured Value Reading
Vibration
Gas path leakage
Loss of sensitivity
EL6010-Uras14:
Uneven emitter
modulation
48
• Disconnect the analyzer unit from the gas preparation system.
• Blow out the gas lines with compressed air or clear them mechanically.
• Change the filter elements and packings.
• Check gas line seal integrity.
• Disconnect the analyzer unit from the gas preparation system.
• Check the analyzer unit gas lines for crimping or loose connections.
• Check the integrity of the analyzer unit gas paths (see page 44 for instructions).
• The outlet pressure should be equal to atmospheric pressure.
• Exhaust gases should be routed to the atmosphere or to an exhaust pipe
directly or via the shortest possible large-diameter line.
• Do not route exhaust gases via flow reducers or shutoff valves.
• Take measures to reduce vibration.
• Increase the time constant (see page 28).
• Check the integrity of the analyzer unit gas paths (see page 44 for instructions).
• Check sample component drift reading (status messages 5 and 6, see page 46).
• Have the emitter and modulator assembly checked by the service department.
EL6010 Gas Analyzer in Category 2G Operator's Manual
42/24-30 EN Rev. 1