Calibrate Sample Component,
Option for
EL6010-Uras14
38
continued
Step
Action
9
Select Span calibration
with test gas.
10
Check the span set
point and adjust as
needed.
11
Turn on the span gas
supply.
12
Start calibration if the
measured value reading
is stable.
13
Calibration in progress.
14
Accept calibration
or
reject calibration.
9
Select Span calibration
with calibration cell.
10
The calibration cell set
point is displayed.
11
Connect the zero gas (if
this has not already been
done).
12
Start calibration if the
measured value reading
is stable.
13
Calibration in progress.
14
Accept calibration
or
reject calibration.
This display shows that an error occurred
during calibration and that calibration
was unsuccessful.
In this case, press Enter to return to the
start of zero or span calibration, check the
test gas concentration, correct as needed
and repeat the calibration process.
EL6010 Gas Analyzer in Category 2G Operator's Manual
Key
Display
Span calibration
Span setpoint
10.00
Vol% CO
Connect span gas!
Meas. value stable?
9.91 Vol% CO
Wait cal. is running
Meas. value ok?
10.00 Vol% CO
Span with cal. cell
Setpoint cal. cell
8.52 Vol% CO
Connect zero gas!
Meas. value stable?
8.49 Vol% CO
Wait cal. is running
Meas. value ok?
8.52 Vol% CO
Amplification error
0.00 Vol% CO
Key
Enter
CO
Enter
Enter
Enter
Enter
Back
Enter
CO
Enter
Enter
Enter
Enter
Back