ABB EL6010 Operator's Manual page 27

Continuous gas analyzers
Table of Contents

Advertisement

Check the Installation,
42/24-30 EN Rev. 1
continued
Check
Connection of electrical lines
• Does the power line voltage agree overall with the permitted operating
voltage (85 to 250 VAC, see rating plate)?
• Are all the electrical lines installed permanently according to the specifi-
cations and correctly connected to the terminal strip in the cable com-
partment?
• Are no loose core ends present? Are all unused core ends insulated and
mechanically secured?
• Are the correct cable types being used for the lines which are led to the
control unit through cable connectors (outside diameter 9 to 10.5 mm)?
• Are the lines firmly seated in the cable glands?
• Are the shielded lines led through the metallic cable connector? Is the
shield braid correctly installed at the cable connector?
• Are the 24-VDC connecting cable and the data transmission cable,
which are non-detachably attached to the analyzer unit, not shortened
to a length of less than 1 meter and not damaged?
Housing integrity
• Is the analyzer unit housing intact?
• Are all flame barriers and screw plugs present?
• If the analyzer units EL6010-Uras14 and -Caldos17 are mounted horizon-
tally, are the O-rings, which are placed between the housing floor and
the housing as well as between the housing and the housing lid in the
slots provided, cleanly laid and not pinched?
• Are all the parts of the housing screwed together fully to the respective
stops and secured from turning with the hex socket head screws?
• Is the control unit housing intact?
• Is the control unit housing firmly closed?
• Is the housing cover screwed on fully to the stop and secured from
turning with the hex socket head screw?
• Is the seal in the cover of the cable compartment intact? Is the cover of
the cable compartment firmly closed?
• Are all cable connectors present and firmly screwed in?
• Are all the openings in unused cable connectors sealed tightly with
sealing plugs?
Connection of peripheral devices
• Are all devices needed for gas conditioning, calibration and exhaust
processing correctly connected and ready for use?
EL6010 Gas Analyzer in Category 2G Operator's Manual
25