Chapter 3

Readout

Screen
Display in
Measuring Mode
Number of Places
after the Decimal
Point
Blinking Screen
42/24-30 EN Rev. 1

Gas Analyzer Operation

The lighted LCD screen consists of two lines of 20 characters each.
CO
105.2 ppm
CO
21.14 %Span
All screen displays shown in this operator's manual are examples. As a rule, actual
displays on the device will differ.
When the screen displays the measured value in physical units (e.g. ppm) the
number of places after the decimal point depends on the size of the measuring
span:
Span (in physical units)
Places
For the display of the measured value as a percentage of the whole measuring
range (%Span) two places are always shown after the decimal point.
The number of decimal places when setting the parameters is the same as in the
display in measuring mode.
Blinking of part of the screen in measuring mode indicates that the gas analyzer or
process system has experienced a fault.
See Chapter "Status Messages, Troubleshooting", page 45, for details on faults,
causes and corrective actions.
EL6010 Gas Analyzer in Category 2G Operator's Manual
In measuring mode the screen displays the sample
component name(s), the measured value in digits
and the physical unit for the measured value.
Press the
key to display the measured value as a
percentage of the measuring range (span).
< 5 ≥ 5...< 50 ≥ 50...< 500 ≥ 500
3
2
1
0
21