The sample gas dew point should be at least 5 °C below the lowest ambient
temperature throughout the sample gas path. Otherwise a sample gas cooler or
condensate trap is required. Water vapor content variations cause volume errors.
The EL6010 gas analyzers must not be used for the measurement of corrosive
gases. Gases like chlorine (Cl
gases or aerosols containing chlorine must be cooled or undergo prior absorption.
The outlet pressure must not exceed the inlet pressure by more than the pressure
loss across the analyzer.
Quantity
Description
1
EL6010-CU Control Unit
1
EL6010-... Analyzer Unit with attached connection cables for
24 VDC power supply and data transmission
1
Analyzer Data Sheet
1
Operator's Manual 42/24-30 EN
1
2.5 mm Hex Wrench
1
220 mm diam. x 3 mm O-Ring
The analyzer data sheet contains the following information:
• Order number (A-No.)
• Part number (P-No.)
• Serial number (F-No.)
• Production date
• Power supply (voltage, frequency, consumption)
• Software version
• Sample components and measuring ranges
• Serial numbers of installed components
The analyzer data sheet must be kept on the gas analyzer so that it is always
ready-to-hand.