Unpack and Install the Gas Analyzer,
EL6010-CU Control
Unit Installation
EL6010-... Analyzer
Unit Installation
EL6010-Magnos106
EL6010-Caldos17
EL6010-Uras14
14
4 M8 or M10 bolts (not supplied) are needed to install the EL6010-CU control unit.
4 M8 bolts (not supplied) are needed to install the EL6010-... analyzer unit.
The EL6010-Magnos106 analyzer unit must be installed with the housing in the
vertical position and the connection ports facing downward.
The EL6010-Caldos17 and EL6010-Uras14 analyzer units can be installed with the
housing in the vertical or horizontal position.
• If installed vertically the connection ports must face downward.
• For horizontal installation the O-ring supplied (∅ 220 x 3 mm) must be installed in
the slot provided between the housing floor and the housing in order to ensure
IP54 housing protection type.
EL6010 Gas Analyzer in Category 2G Operator's Manual
continued