EL6010-Magnos106: Measuring Range

Measuring Range
Step Interval
Reference
Measuring Range
Measuring Range
Notes
30
Measuring range
0.00 - 25.00 Vol%
Measuring range
0.00
- 25.00 Vol%
Measuring range
0.00 -
25.00
Vol%
The measuring range adjustment step interval is 0.5 Vol.-%.
Calibration of the measuring range should be verified after changing measuring
range limits.
The gas analyzer has a physical measuring range ("reference measuring range").
The limits of this range are 0 Vol.-% O
One measuring range can be selected within the limits of the reference measuring
range. It is factory-set and calibrated to 0 to 25 Vol.-% O
The smallest measuring range is 0 to 1 Vol.-% O
range is 0 to 100 Vol.-% O
Measuring ranges within ignition limits should not be attempted. The gas inlet
conditions should be observed (see page 9).
EL6010 Gas Analyzer in Category 2G Operator's Manual
Measuring range
Measuring range start
Value range: Within the reference measuring range
Measuring range end
Value range: Within the reference measuring range
and 100 Vol.-% O
2
; the largest adjustable measuring
2
.
2
.
2
.
2
42/24-30 EN Rev. 1