Technical Data - M&C PSS Series Instruction Manual

Portable gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

9

Technical data

Gas Conditioning Type
Part No.
Sample gas outlet dew point
Sample gas outlet dew point
stability
Sample gas inlet temperature
Sample gas inlet water vapor
saturation
Heat exchanger gas flow rate
Ambient temperature
Storage temperature
Pressure
Total cooling power*
Number of gas inlets
Number of gas outlets
Medium connections
Material of parts beeing in
contact with the medium
Ready for operation
Mains power supply
Power consumption
Fuse protection
Electrical connection
Electrical equipment standard
Case protection
Housing
Housing color
Housing dimensions
(W x H x D)
Weight
PTFE=Polytetrafluoroethylene (Teflon®), PVDF=Polyvinylidenfluoride
* Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C [77 °F] ambient
temperature and an outlet dew point of 5 °C [41 °F].
www.mc-techgroup.com
PSS-5C
01G4000
Range of adjustment: +2 to 15 °C [35.6 to 59 °F], factory setting:
+5 °C [41 °F]
At const. conditions: < ±0.1°C [±0.18 °F]
*Max. 80°C [176 °F], optional: *max. 180°C [356 °F] with stainless
steel bulkhead union
*Max. +80 °C [176 °F]
*Max. 150 Nl/h
*5 to 40 °C [41 to 104 °F]
-25 to +65 °C [-13 to +149 °F]
0.7 bar to 1.4 bar abs.
Max. 80 kJ/h
1
1, optional: max. 2
Tube connections 4/6 mm
Stainless steel, glass, PVDF, PTFE, Novoprene
Approx. 3 min. (with heated line approx. 1 h)
115 or 230 V AC ±10 % 50/60 Hz
Max. 240 VA; with option temperature controller and heated
sample line: 230 V, max. 1620 VA; 115 V, max. 920 VA
4 A t, 5 mm x 20 mm; with option temperature controller: 10 A t,
5 mm x 20 mm
2 m [≈ 6.56 ft] length cable
EN 61010
IP42 EN 60529, with upright case and door closed. Unused
connectors must be closed with appropriate caps.
Portable impact-resistant case with integrated trolley system and
pull-out handle
Black
451 x 654 x 279 mm [≈ 17.8" x 25.7" x 11"] with casters and handle
Approx. 17 kg [≈ 37.5 lbs]
PSS-5C | 1.02.02
PSS-5/2C
01G4250
*Max. 150 Nl/h per
heat exchanger
PSS-5/3C
01G4500
*Max. 350 Nl/h
®
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pss-5cPss-5/2cPss-5/3c