Figure 8 Case Content: Pss-5C And Pss-5/3C - M&C PSS Series Instruction Manual

Portable gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

1
Case door with display
2
Case
3
Terminal mounting rail
4
Fine filter FP-2T
Figure 8
Case content: PSS-5C and PSS-5/3C
All components of the gas conditioning system are freely accessible and built into a portable case 2.
The case door
1
can be opened to the left by loosening the latches mounted on the side and top of the
case.
The installation of the gas cooler
required maximum gas volume flow.
The minimum flow rate is determined by the sample gas pump 5. If the required minimum total flow rate
is not reached, excessive overpressure can lead to premature destruction of the pump diaphragm. The
maximum pressure is reduced by using an optional needle valve in the bypass of the sample gas pump
(Part No. 01G9050).
The gas cooler is equipped with a Duran glass heat exchanger
or stainless steel are optionally available.
The FP-2T fine filter (2 µm filter porosity) 4, installed upstream of the sample gas pump 5, ensures the
necessary solids separation.
The overtemperature alarm contact (+8 °C [46.4 °F]) of the cooler automatically regulates the
switching on and off of the sample gas pump 5.
The resulting condensate is continuously discharged by a peristaltic pump type SR25.2-W 6.
www.mc-techgroup.com
1
2
5
Sample gas pump
6
Peristaltic pump
7
Gas coolerr
8
Heat exchanger
7
and a corresponding diaphragm sample gas pump
PSS-5C | 1.02.02
8
as standard. Heat exchangers in PVDF
3
4
5
6
7
8
5
depends on the
19