Table of Parameter Codes
No.
Description:
1
Software version
5
Brightness setting
of display
10
LA on/off switching 0
11
Sensitivity LA
12
HIGH dT
13
LOW dT
14
Hysteresis
(temperature
alarm)
15
mA
range
selection
70
Operating time in
days
84
Fan speed
210
Calibration LA dry
211
Calibration LA wet
212
Calibration
open / cable break
213
Display LA dry
214
Display LA wet
215
Display LA open
310
mA output LOW
Calibration
311
mA output HIGH
Calibration
48
Default: Range:
5
0 - 9
0,1,2
2
1-7
3
2 - 8
[°C]
3
2 - 8
[°C]
2
1,2 [°C]
4-20
0-20 / 4-
20 [mA]
1
0-5
LA
1 mA
20 mA
PSS-5C | 1.02.02
Note:
Brightest display setting is 9
0=off; 1= without cable break detection; 2= with
cable break detection
The higher the value is, the sooner the alarm is
triggered.
Differential temperature between set point and
upper temperature alarm limit
Differential temperature between setpoint and
lower temperature alarm limit
As soon as a temperature alarm limit of "2" is
selected, the hysteresis is automatically reduced to
1
Corresponds to the temperature range: -10 °C to
+50 °C (4 – 20 mA: at 5 °C = 8 mA)
This value will not be deleted even after restarting
Value has no influence on the output power
Connect the sensor and confirm with the
Hold the sensor in water and confirm with the
key.
Disconnect the sensor and confirm with the
The cable break detection triggers in the middle
between this value and that of dry.
Check here the current value
Check here the current value
Check here the current value
Connect the multimeter and press or hold
-keys until the display matches 1mA.
Connect the multimeter and press and hold
-keys until the display matches 20 mA.