The 4/6 mm tube connections for the condensate and sample gas lines are located on the side of the
case.
A stainless steel sample probe (length 0.5 m, Ø 6 mm) and a 3 m PVC sample tube (4/6 mm) are
included in the standard scope of delivery.
The ventilation grids in the door and in the left side wall of the case provide sufficient convex forced
ventilation.
Options:
The PSS-5... sample gas conditioning unit can be equipped at the factory with a maximum of two
sample gas outlets. Each sample gas outlet can be controlled by the optional installation of a flow
meter type FM40 with needle valve according to the specified volume flow range (see the option table
in chapter '9 Technical data').
Unused mounting holes for sample gas outlets or flow meters are closed with blind caps.
I To protect the downstream analyzers against liquid ingress and to increase the operational reliability
of the entire system, we recommend the installation of a liquid alarm sensor type LA1S (Part No.
01G9015). For this purpose, the FP-2T fine filter installed as standard is replaced by the FP-2T-D fine
filter with mounted liquid alarm sensor at the factory. The evaluation electronics for the LA1S liquid
sensor are integrated as standard in the ECPX000C cooler.
The PSS-5... can also be equipped with a special sample gas inlet (Part No. 01G9060) for connecting
a heated line. The existing kink protection can only be used for heated lines with connection type "C"
(Part No. 03B1012). Assembly instructions can be found in chapter "18.1.1 Connecting the heated
sample line with special adapter (option)".
It is also possible to connect the heated line (Part No. 01B4036) in connection with the gas sampling
probe PSP 4000.
The temperature controller 701 (Part No. 01G9055) which is necessary for the regulation of the heated
line is mounted on the terminal rail
3
(Figure 8) at the factory with the required option.
A 3-way ball valve (Part No. 01G9046) or a 5-way ball valve (Part No. 01G9045) can optionally be
installed in the inlet of the gas conditioning unit for test gas feeding or sample gas switching.
20
PSS-5C | 1.02.02