7
TECHNICAL DATA
Gas Conditioning Unit Series PSS
Part No.
Gas outlet dew point
Gas outlet dew point stability
Gas inlet temperature
Gas inlet water vapor saturation
Gas flow rate
Ambient temperature
Storage temperature
Pressure
Total cooling capacity
Number of gas inlets
Number of gas outlets
Medium connections
Material of sample contacting parts
Ready for operation
Mains power supply
Power consumption
Liquid detection
Fuse protection
Electrical connection
Housing
Case protection
Housing color
Housing dimensions (H x W x D)
Weight
10
PSS-20
®
760-280200000
Range of adjustment: +2 to +15 °C [35.6 to 59 °F], factory
setting: +5 °C [41 °F]
At const. conditions: < ±0.1 °C [±0.18 °F]
*Max. 180 °C [356 °F] with stainless steel bulkhead union
*Max. +80 °C [176 °F]
*Max. 150 Nl/h
+5 to +40 °C [41 to 104 °F]
-25 °C to +65 °C [-4 to 149 °F]
0.7 to 1.4 bar abs.
*Max. 90 kJ/h
1
2 [bypass]
Tube connections DN 4/6 mm
Stainless steel 316Ti, glass, PVDF, PTFE, Novoprene, FKM
Approx. 5 min.
230 V AC ±10 % 50/60 Hz
Max. 400 VA, with option temperature controller and
heated sample line: 230 V max. 1380 VA
Yes
4 A, time-delay, 5 x 20 mm (with option temperature
controller: 10 A, time-delay)
2 m [≈ 6.6 ft] long universal power cord three-hole national
standard plug
Impact-resistant case with handle
IP20
Black/white
Approx. 245 x 341.5 x 411 mm [≈ 9.6" x 13.4" x 16.2"]
without anti-collision angles
Approx. 12.5 kg [≈ 27.6 lbs]
PSS-20 | 1.00.01