Storage Precautions

If the product is to be stored for a long period of time after
delivery, observe the following points.
 The product should be stored in its original packing
condition in the storage location. When the PTFE
lining is specified, the particle board is attached to the
flowmeter. Keep the particle board attached until the
flowmeter is install to the pipe.
 Select a storage location that fulfils the following
conditions:
• A place where it will not be exposed to rain or water
• A place subject to minimal vibrations or shocks
• Temperature and humidity levels should be as follows:
Temperature: -10 to 70°C
Humidity: 5 to 80% RH (no condensation)
The preferred ambient temperature and humidity
levels are 25°C and approximately 65% RH.
 If the product is transferred to the installation site and
stored without being installed, its performance may
be impaired due to the infiltration of rainwater and
so forth. Be sure to install and wire the product as
soon as possible after transferring it to the installation
location.