YOKOGAWA AXG1A Installation Manual page 68

Magnetic flowmeter, size: 25 to 400 mm 1 to 16 in.
Hide thumbs Also See for AXG1A:
Table of Contents

Advertisement

 Warning
The device works normally and measurement is also normal but warning occurs.
Error Message
NE107
Status
Display
S
080:AO 1 saturate
Saturation of Analog output 1 was detected.
S
081:AO 2 saturate
Saturation of Analog output 2 was detected.
S
082:Pls 1 saturate
Saturation of Pulse output 1 was detected.
S
083:Pls 2 saturate
Saturation of Pulse output 2 was detected.
S
084:AI saturate
Saturation of Analog input was detected.
C
085:Cable miscon
Misconnection of cable was detected.
C
086:Coil insulation
Insulation deterioration of coil was detected.
The resistance value of electrode exceeded Level 3.
M
087:Adhesion lv 3
(Adhesion detection of insulation to electrode)
N
088:LC warn
Decrease of conductivity was detected.
M
089:Insu detect
Insulation deterioration of electrode was detected.
Result of Autozero adjustment |Mag Flow Zero| exceeded
C
092:AZ warn
10 cm/s.
C
093:Verif warn
Interruption of verification function was detected.
C
094:Fact noise warn
Fluctuation of flow became larger.
Test mode was executed for any of Flow velocity,
C
095:Simulate active
Volumetric flow rate, Mass flow rate, Calorie, Current
output, Pulse output, Status input, Status output.
S
096:AO 1 fix
It was detected that fixied value is set to Current output 1.
S
097:AO 2 fix
It was detected that fixied value is set to Current output 2.
S
098:Pls 1 fix
It was detected that fixied value is set to Pulse output 1.
S
099:Pls 2 fix
It was detected that fixied value is set to Pulse output 2.
S
100:AI fix
It was detected that fixied value is set to Current input.
C
101:Param restore run
Restore function of parameter is running.
N
102:Disp over
Number of digits available for display exceeded the limit.
N
103:SD size warn
Free space of microSD card decreased to less than 10%.
M
104:Bkup incmplt
Parameter backup failed.
S
105:SD mismatch
Mismatch of microSD card was detected.
M
106:SD removal ERR
Removal of microSD card failed.
N
131:Trans mismatch
Mismatch of sensor and transmitter was detected.
 Information
The device works normally and measurement is also normal. Just reference information.
Error Message
NE107
Status
Display
N
120:Watchdog
Error of Watchdog timer was detected.
N
121:Power off
Power-off was detected.
N
122:Inst power FAIL
Instantaneous power failure was detected.
N
123:Param bkup run
Parameter backup is running.
N
124:Data log run
Data log is running.
N
130:DevID not enter
No data entry of Device ID was detected.
The Modbus configuration has been changed, but the
N
132:Modbus cfgupdate
device is not operating according to its configuration.
As a result of self-test, inconsistency of internal circuit was
N
133:G/A mismatch
detected.
N
246:Simulate SW OFF
Writing to parameters is locked.
N
248:Simulate SW ON
Software or hardware simulation switch is ON.
<7. Errors and Countermeasures (Display unit)>
Error Description
Error Description
Countermeasure
Check process value and current output 1 parameter setting.
Check process value and current output 2 parameter setting.
Check process value and pulse output 1 parameter setting.
Check process value and pulse output 2 parameter setting.
Check current input value and current input parameter setting.
Check the signal cable and excitation cable connection.
Contact Yokogawa service center.
Recommend cleaning electrode.
Check fluid conductivity.
Contact Yokogawa service center.
Check fluid is stopped when executing auto-zero adjustment.
Execute Verification again.
Check the fluid.
Release simulation or test mode.
Check current output 1 is test mode or not.
Check current output 2 is test mode or not.
Check pulse output 1 is test mode or not.
Check pulse output 2 is test mode or not.
Check current input is test mode or not.
Check Display format parameter.
microSD card may run out of memory space.
Retry parameter backup.
Change microSD card.
Remove microSD in appropriate rocedure.
Contact Yokogawa service center.
Countermeasure
Contact Yokogawa service center.
Contact Yokogawa service center.
Restart Modbus communication.
Contact Yokogawa service center.
Change the RB Write Lock (RB. WRITE_LOCK) to Not Locked or
turn off the hardware write lock switch, depending on RB Feature
Selection (RB.FEATURE_SEL).
Delete the value of MTB Sim Enable Message (MTB.SIM_ENABLE_
MSG) or turn off the hardware simulation switch.
67
7
IM 01E24A01-01EN

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Axw4aAxw seriesAdmag ti seriesAxfa11gAx01c

Table of Contents