 Setting Alarm
The device works normally but parameter setting error occurs.
Parameter setting is needed.
Error Message
NE107
Status
Display
Setting error of flow span was detected. (fulfill "0.05 m/s <
S
060:Span cfg ERR
Span < 16 m/s")
S
061:PV F cfg ERR
PV was set to flow noise.
Setting error of Current output 1 was detected. (fulfill "LRV
S
062:AO 1 4-20 lmt
< HRV")
Setting error of Current output 2 was detected. (fulfill "LRV
S
063:AO 2 4-20 lmt
< HRV")
Setting error of ararm high/low limit function was detected.
N
065:H/L cfg ERR
(fulfill "Hi Alarm - Lo Alarm > H/L Hys" and "HH Alarm - LL
Alarm > HH/LL Hys")
Setting error of density value was detected when PV was
S
066:Density cfg ERR
set to mass flow rate.
S
067:Pls 1 cfg ERR
Setting error of Pulse output 1 was detected.
S
068:Pls 2 cfg ERR
Setting error of Pulse output 2 was detected.
Configuration error of nominal size was detected. (fulfill
C
069:Nomi size cfg
"0.99 mm < nominal size < 3000.10 mm (0.01 inch <
nominal size < 120.10 inch)")
Setting error of electrode adhesion detection function
C
070:Adh cfg ERR
wasdetected. (fulfill "Level1 < Level2 < Level3 < Level4")
Setting error of flow noise verification function wasdetected.
C
071:FLN cfg ERR
(fulfill "Level1 < Level2 < Level3 < Level4")
C
072:Log not start
Data logging failed to start.
C
247:RB O/S Mode
Resource Block is in O/S mode.
C
300:AI1 O/S Mode
AI1 Block is in O/S mode.
C
301:AI1 Man Mode
AI1 Block is in Man mode.
C
302:AI1 Not Schedule
AI1 Block is not scheduled.
C
303:AI1 Simulate Act
AI1 block is in simulation mode.
C
304:AI2 O/S Mode
AI2 Block is in O/S mode.
C
305:AI2 Man Mode
AI2 Block is in Man mode.
C
306:AI2 Not Schedule
AI2 Block is not scheduled.
C
307:AI2 Simulate Act
AI2 block is in simulation mode.
C
308:AI3 O/S Mode
AI3 Block is in O/S mode.
C
309:AI3 Man Mode
AI3 Block is in Man mode.
C
310:AI3 Not Schedule
AI3 Block is not scheduled.
C
311:AI3 Simulate Act
AI3 block is in simulation mode.
C
328:PID O/S Mode
PID Block is in O/S mode.
C
329:PID Man Mode
PID Block is in Man mode.
C
330:PID Not Schedule
PID Block is not scheduled.
C
331:PID Bypass Mode
PID Block is in Bypass mode.
C
332:IT1 O/S mode
IT1 Block is in O/S mode.
C
333:IT1 Man mode
IT1 Block is in Man mode.
C
334:IT1 Not Schedule
IT1 Block is not scheduled.
C
336:IT2 O/S Mode
IT2 Block is in O/S mode.
C
337:IT2 Man Mode
IT2 Block is in Man mode.
<7. Errors and Countermeasures (Display unit)>
Error Description
Countermeasure
Check or change span parameter setting.
Check or change PV FLOW SEL parameter setting.
Check or change Analog output 1 parameter setting.
Check or change Analog output 2 parameter setting.
Check and change H/L, HH/LL parameter setting.
Check and change Density parameter setting.
Change Pulse output 1 parameter setting.
Check and change Pulse output 2 parameter setting.
Check and change nominal paramter setting.
Check and change adhesion parameter setting.
Change flow noise paramter setting.
Insert microSD card.
Change the RB Block Mode.Target (RB.MODE_BLK.Target) to
Automode.
Change the AI1 Block Mode.Target (AI1.MODE_BLK.Target) to Auto
or other mode.
Change the AI1 Block Mode.Target (AI1.MODE_BLK.Target) to Auto
or other mode.
ScheduleAI1 Block.
Change the AI1 Simulation En/Disable (AI1.SIMULATE.SIMULATE_
ENABLE) to Disabled.
Change the AI2 Block Mode.Target (AI2.MODE_BLK.Target) to Auto
or other mode.
Change the AI2 Block Mode.Target (AI2.MODE_BLK.Target) to Auto
or other mode.
ScheduleAI2 Block.
Change the AI2 Simulation En/Disable (AI2.SIMULATE.SIMULATE_
ENABLE) to Disabled.
Change the AI3 Block Mode.Target (AI3.MODE_BLK.Target) to Auto
or other mode.
Change the AI3 Block Mode.Target (AI3.MODE_BLK.Target) to Auto
or other mode.
ScheduleAI3 Block.
Change the AI3 Simulation En/Disable (AI3.SIMULATE.SIMULATE_
ENABLE) to Disabled.
Change the PID Block Mode.Target (PID.MODE_BLK.Target) to Auto
or other mode.
Change the PID Block Mode.Target (PID.MODE_BLK.Target) to Auto
or other mode.
Schedule PID Block.
Change the parameter of Bypass (PID.BYPASS) to OFF.
Change the IT1 Block Mode.Target (IT1.MODE_BLK.Target) to Auto
or other mode.
Change the IT1 Block Mode.Target (IT1.MODE_BLK.Target) to Auto
or other mode.
Schedule IT1 Block.
Change the IT2 Block Mode.Target (IT2.MODE_BLK.Target) to Auto
or other mode.
Change the IT2 Block Mode.Target (IT2.MODE_BLK.Target) to Auto
or other mode.