Electrical Requirements - Labconco Protector XStream 11041 Series User Manual

Laboratory fume hoods
Table of Contents

Advertisement

Original instructions
Projected Airflows and Static Pressure
Hood Type
Face
Sash at
Full open
at 28"
XSTREAM
100
80
60
50
40
Hood Type
Face
Sash at
62.50%
open at 18"
XSTREAM
100
80
60
Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a remote blower system.

Electrical Requirements

The Protector XStream Hood models feature internal wiring for the LED light
assembly and light switch. All internal wiring is terminated at the single point
wiring junction box for hook-up by a qualified electrician. The blower switch,
and light switch wires are also terminated at the single point wiring junction box
for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C: Protector XStream Specifications for the wiring diagram for proper
electrical installation.
Les modèles Protector XStream Hood sont dotés d'un câblage interne pour
l'éclairage LED et l'interrupteur. Tout le câblage interne est terminé à la boîte de
jonction point de câblage unique pour le raccordement par un électricien qualifié.
Le bouton du ventilateur, et fils de l'interrupteur de lumière sont également mis
fin à la boîte de jonction point de câblage unique pour le raccordement par un
électricien qualifié. Reportez-vous au Chapitre 3: Mise en route et à l'Annexe C:
Spécifications Premier Protector pour le schéma de câblage pour l'installation
électrique correcte.
Product Service 1-800-522-7658
Eco Energy Saving Air Foil
Velocity (fpm)
Airflow Volumetric Rate (CFM) @
Sash at 62.5%
Static Pressure (inches of water)
open
4'
5'
at 18"
Hood
Hood
160
705, 0.26"
930, 0.32" 1150, 0.41" 1600, 0.29"
128
565, 0.17"
745, 0.20"
96
425, 0.09"
560, 0.12"
80
350, 0.06"
465, 0.08"
64
280, 0.04"
370, 0.05"
Eco Energy Saving Air Foil
Velocity (fpm)
Airflow Volumetric Rate (CFM) @
Sash at
Static Pressure (inches of water)
Full open
4'
5'
at 28"
Hood
Hood
62
440, 0.10"
580, 0.12"
50
350, 0.06"
465, 0.08"
38
265, 0.04"
350, 0.05"
Chapter 2: Prerequisites
Flush Air Foil
Airflow Volumetric Rate (CFM) @
Static Pressure (inches of water)
6'
8'
4'
Hood
Hood
Hood
740, 0.28"
920, 0.26" 1280, 0.19"
590, 0.18"
690, 0.15"
960, 0.10"
445, 0.10"
575, 0.10"
800, 0.07"
370, 0.07"
460, 0.07"
640, 0.05"
295, 0.05"
Flush Air Foil
Airflow Volumetric Rate (CFM) @
Static Pressure (inches of water)
6'
8'
4'
Hood
Hood
Hood
720, 0.16" 1000, 0.11"
475, 0.11"
575, 0.10"
800, 0.07"
380, 0.07"
430, 0.06"
600, 0.04"
285, 0.04"
5'
6'
8'
Hood
Hood
Hood
975, 0.34" 1205, 0.44" 1675, 0.31"
780, 0.22"
965, 0.28" 1340, 0.20"
585, 0.12"
725, 0.16" 1005, 0.11"
490, 0.08"
605, 0.11"
840, 0.08"
390, 0.06"
480, 0.07"
670, 0.05"
5'
6'
8'
Hood
Hood
Hood
625, 0.13"
775, 0.17" 1075, 0.12"
500, 0.08"
620, 0.11"
860, 0.08"
375, 0.05"
465, 0.06"
645, 0.04"
7