Chapter 6: Maintaining Your Protector XStream
 The hood baffles should be checked for blockages behind them to ensure
 Les déflecteurs de hotte devraient être contrôlés pour leur blocage arrière
 All weekly activities.
 Toutes les activités hebdomadaires.
Annually
Annuellement
 Replace the lamps if needed.
 Remplacer les lampes, si nécessaire.
 Have the fume hood recertified by a qualified certification technician. See
 Recertifier la hotte aspirante par un technicien agréé en certification.Voir
 All monthly activities.
 Toutes les activités mensuelles.
Biannually
 The sash assembly should be checked to ensure proper operation and to

Routine Service Operations

Front Panel Removal:
1. Simply lift the front panel up and then away from the hood to provide access
to the top.
Changing the Lamp:
S
A
1. Turn light switch to "OFF".
2. Remove the front panel as noted earlier.
3. Reach over the front header of the hood and remove the knock out plugs at
both ends of fixture. Lift fixture up.
4. Remove the lamp by pushing it out of the spring-loaded lamp socket and
swinging it out of the other lamp socket.
5. Install the new lamp by reversing the removal procedure.
32
that the hood is maintaining proper airflow.
afin d'assurer que la hotte maintient un flux d'air correct.
"Certifying the Protector Fume Hood" in Chapter 3.
au Chapitre 3 « Certifier la Hotte Aspirante Protecteur ».
make sure there are no signs of abnormal wear on the sash pulleys, cables
and clamps.