Labconco Protector XStream 11041 Series User Manual page 45

Laboratory fume hoods
Table of Contents

Advertisement

Original instructions
PROBLEM
Electrical duplex
outlets no longer
have power
Service valves no
longer operate
Product Service 1-800-522-7658
CAUSE
Vertical sash frame is
distorted.
Cable is frayed or has
slipped off the
pulleys.
Wires not connected
or faulty duplex.
Circuit breakers
tripped in building
electrical supply.
Faulty building
supply.
Valve no longer
operates.
Supply line or outlet
line has leaks.
Chapter 8: Troubleshooting
CORRECTIVE ACTION
Place horizontal glass symmetrically
and pull sash down to air foil.
Straighten damaged frame.
Re-install, cable must be replaced
immediately if damaged.
Check wire connection or replace
duplex.
Reset circuit breakers.
Inspect building supply shut off valves
and appropriate pressures below the
recommended 40 PSI.
Replace valve and check for leaks.
Inspect line for leaks and fix any
leaking plumbing connections.
41