11.5.4. Set Foldback Protection
This command enables or disables the foldback protection. When the FB protection is
enabled, if the power supply goes into constant-current (CC) mode for about a ½
second, then the output will turn off, the front panel will show "Fb" and the
Questionable Condition register bit 3 will be set high.
This protection is normally used to prevent a hazard condition if the output load goes
into a short-circuit. For applications where the power supply is normally operating in
CC mode, such as driving electro-magnets, the FB protection cannot be used.
The FB setting is seen on the supply's front panel FOLD LED. A foldback fault is
cleared by turning the power supply output on again.
Syntax:
Parameter: 0 or OFF will turn the foldback protection OFF
Example:
Query:
11.5.5. Read Foldback Tripped
This query reports if a foldback protection fault has occurred. If a fault has occurred, it
may be cleared by turning the power supply output on again (see the
OUTP:STAT ON command, section 11.2.3)
Syntax:
Example:
Responses:

11.6. LAN Specific Commands

11.6.1. Blinking the Identify LED
When communicating to a stack of Genesys™ power supplies, it may be difficult to
know which supply is selected for communicating. The LAN computer may send a
command to cause the selected supply's front panel displays to blink on and off. Also,
the LAN LED on the rear panel will blink. In a multi-drop chain of supplies (see
section 12), only the LAN supply will blink.
If blinking is turned on, then it may be turned off by sending the OFF command or by
changing any front panel control.
Syntax:
Example:
Query:
8303460000 Rev A
[SOURce:]CURRent:PROTection:STATe <0|1|OFF|ON>
1 or ON will turn the foldback protection ON
CURR:PROT:STAT 1 (or CURR:PROT:STAT ON)
CURR:PROT:STAT? will return ON in the example, otherwise OFF
[SOURce:]CURRent:PROTection:TRIPped?
CURR:PROT:TRIP?
0 when no FB fault is active
1 when supply is shut-down because of a FB fault
SYSTem:COMMunicate:LAN:IDLED <0|1|OFF|ON>
SYST:COMM:LAN:IDLED ON
There is no query to read if flashing is turned on.
43