SYSTEM:ERROR? MESSAGES

These are the system error messages which may be returned from the "SYST:ERR?" query.
See section 11.8.1 for a description of this system error queue
ERROR
ERROR MESSAGE
"No error"
0
"Command error"
-100
"Invalid Character"
-101
"Syntax error"
-102
"Data type error"
-104
"Missing parameter"
-109
"Program word too long"
-112
"Invalid Suffix"
-131
"Data out of range"
-222
"Hardware Missing"
-241
"Queue Overflow"
-350
"Execution error"
+300
"PV above OVP"
+301
"PV below UVL"
+302
"OVP below PV"
+304
"UVL above PV"
+306
"On during fault"
+307
"Fault shutdown"
+320
"AC fault shutdown"
+321
"Over-Temperature"
+322
"Fold-Back shutdown"
+323
"Over-Voltage shutdown"
+324
"Analog shut-off shutdown"
+325
"Output-Off shutdown"
+326
"Enable Open shutdown"
+327
8303460000 Rev A
ERROR DESCRIPTION
No error reported
LAN card receives command with unspecified error
A character was received that is not:
a-z, A-Z, 0-9, ?, *, :, ;,period, space, CR, LF
LAN card receives unrecognized command word
LAN card receives command parameter with wrong type of data.
Example: a letter where a number is expected.
Valid command received but not enough parameters.
Command word had more than 14 characters before separator
(space or colon) was found
Attempt to select multi-drop address > 30
Attempt to program voltage, current or OVP beyond supply limits
When operating in RS-485 Multi-drop mode, an attempt was made
to address a non-existent supply.
Too many SYST:ERR messages are stored in this queue and the
newest messages are discarded. The max. number of stored
messages is 10.
General execution error.
Attempt to program voltage above OVP setting.
(or more exactly, PV not "5% of voltage rating" below OVP)
Attempt to program voltage below UVL setting.
(or more exactly, PV not "5% of voltage rating" above UVL)
Attempt to set OVP below voltage setting.
(or more exactly, OVP not "5% of voltage rating" above PV)
Attempt to set UVL above voltage setting.
(or more exactly, UVL not "5% of voltage rating" below PV)
Attempt to set supply output "ON" when a fault exists
General message for non-specified shutdown
Brown-out or phase-loss shutdown occurred
Over-temperature shutdown occurred
Fold-Back shutdown occurred
Over-Voltage shutdown occurred
Shut-Off occurred from rear panel J1
Output-Off occurred from front panel button
Enable Open occurred from rear panel J1
61