USING THE ERROR AND STATUS REGISTERS
The SCPI error and status registers report many conditions, statuses and events into
short numeric codes. Momentary events may be latched so the remote computer can
read about them at a later time. A "fan out" architecture is used to summarize all the
data into one Status Byte register.
The Genesys™ power supply, with the LAN option, does not support
the service request event defined in the Status Byte register, bit 6.
There are three types of registers. Data, as in a bit is set high or low, flows top to
bottom in the following list. See Figure 7 for the diagram.
Condition Registers: contain bits that are set when a condition or error
occurs. The bits are only cleared when the condition or error is cleared. The
contents may be read but not changed.
Enable Registers: individual bits may be set to allow the status and errors to
be latched for reading even after the status or error condition has cleared. The
power-up default is all zeroes, that is, no latching occurs.
Event Registers: bits that are set when an event or error occurs. The bits are
cleared when the contents of the register are read or when a clear status
command is sent.
11.8.5. Clear Status
This command resets the error, status and message registers in the LAN board. No
power supply settings are changed.
The *CLS command will:
 Erase any response message in the output queue
 Erase any errors in the SYSTEM:ERROR queue
 Reset all event registers to zeroes
 Will not change any Enable or Condition register bits
Syntax:
Query:
8303460000 Rev A
*CLS
This command has no query form
49
NOTE
IA672-04-01-Rev. I