Calibration and Linearity (Continued)
Pipetting errors can be identified by comparing duplicate readings. The instrument should
give the expected value ± (1% of the expected value + .01A).
For example, if the 1/4 dilution reads .520 A, then you can expect your 1/2 dilution sample to
read twice as much or 1.04 A ± [(1% of 1.04) +0.01 A], which is ± .02.
A good working instrument would then give a result between 1.02 A and 1.06 A. You will
know that your filters are deteriorating when darkly colored solutions consistently read
lower than expected.
Since the procedures for verifying the calibration and linearity of a vertical photometer are
quite cumbersome and not conducive to stringent criteria, the best way to assure quality
instrument performance is with DRI-DYE® Check Strips.
3.2.2 Storage
The instrument may be stored under the following recommended environmental conditions:
•
Temperature: 10 to 50°C
Humidity: Less than 80% relative humidity, non-condensing.
•
Store the instrument using original packaging if possible.
Perform the following steps before storing:
Set the power switch to OFF and remove the power cord.
Place the instrument in the original packaging material.
When returning the instrument to service from storage, it is recommended that
functional tests be performed as if setting up the instrument for the first time.
Stat Fax
®
4200 Operator's Manual Doc. 4200 Rev. K
51
Need help?
Do you have a question about the Stat Fax 4200 and is the answer not in the manual?