Checkout Procedure - Awareness Technology Stat Fax 4200 Operator's Manual

Microplate reader
Table of Contents

Advertisement

1.7 Checkout Procedure

Follow this procedure to verify that the instrument is ready for use.
Visually confirm the following items:
Power supply is connected to the unit and the power plug is inserted into an AC
outlet.
Power switch is set to OFF.
The instrument is now ready for power-up. Confirm that the instrument responds as
described.
Set the power switch to the ON position.
The printer will print out the
instrument model, serial
number, laboratory name (if
defined under the Settings
menu Laboratory Name option;
the default is no laboratory
name), firmware version,
revision level, date and time
(Figure 1.7-2).
Click Tray Out and make sure
door opens and microplate tray
extends out.
Click Tray In and make sure
door closes.
Before running any assays,
read at least one plate in
absorbance mode to verify
plate carrier operation
If the instrument still produces
results other than those
described here, refer to Section
4. Troubleshooting, or contact
your dealer for assistance.
Stat Fax
Checkout Procedure
Figure 1.7-1 Power up Screen Display
Stat Fax 4200
Laboratory Name (if defined)
Serial Number: 9999
12-20-2018
Figure 1.7-2 Representation of the power up
print out
®
4200 Operator's Manual Doc. 4200 Rev. K
AWARENESS
TECHNOLOGY
INC
Stat Fax
4200
Loading ...
Ver: Eng-01.26.00
16:34:40
11