The Run Test feature allows the operator to recall user tests that have been stored in the
instrument's memory (see example in Figure 2.3.1-1). The Stat Fax
120 complete test setups in nonvolatile memory, making it easy for the user to recall
complete test configurations.
Select Run Test from the Main display. Tests programmed and saved on the instrument
will display sorted by their test number. Each of the test parameters, including the mode,
wavelengths, standards, units, and the ranges are all stored for reuse.
Blanks and standards (including entire standard curves) that have been read are also
saved. When the test is recalled, the user has the option of using the previous saved
curve or reading a new one.
Figure 2.3.1-1 Run Test - Example of stored test menu
Feature
Arrows
Select
By #
Cancel
14
Stat Fax
1 – TSH
2 – T3
SAMPLE
3 – T3 Uptake
4 – T4
5 – T4 Total
6 – FSH
7 – hCG
8 – LH
9 – Testosterone
Use the ^ UP and
selection; use the >> side arrows to advance to
the next screen; use << arrows for previous
screen
The highlighted test is executed once the Select
key is pressed
Allows the user to enter a specific test number to
recall.
Returns to the main power on display screen
®
4200 Operator's Manual Doc. 4200 Rev. K
®
Cancel
<<
v
>>
Select
By #
Function
DOWN arrow keys to highlight
V
4200 stores up to