Single Standard Mode (Continued)
Selecting any of the Control fields will open a Test Definition Control screen with options to
enable the Control (Yes or No), input the number of Replicates, Name the Control, input a
Low and a High Range, set the Action to Take (Warn, Continue or End Test), input a Lot
Number, input an Expiration Date, and input a Location. Press the Done button when
finished with editing fields.
When Controls are read in duplicate, the mean of the duplicates can be calculated and that
one value used, or, each Control can be required to fall within the ranges individually. The
instrument will prompt 'Check Mean?' once duplicate Controls are selected.
TEST DEFINITION Control 1
When setting the Controls, if there are Replicates, the Location will be the starting location.
If the Controls or their Replicates overlap due to the plate setup, an error flag will display
while running the test to warn the operator.
The Single Standard Mode Test Definition Final page will display and allow editing of
Interpretation Mode, Units, Decimals, and number of Sample Replicates. Use the arrow
keys to advance through the options; press the Edit button to change the values of the
selections on the screen. Press Save to save the selections.
Single Standard Mode Test Definition - Final Page
Stat Fax
Enabled
# Replicates
Control 1
Name
Range Low
Range High
Warn, Continue
Action
Lot Number
Expiration [mm.yyyy]
Location
Test Definition Control screen
TEST DEFINITION Final page
Interpretation Mode
Units
Decimals
# Sample Replicates
<<
Save
®
4200 Operator's Manual Doc. 4200 Rev. K