TABLE OF CONTENTS 1. Introduction ...................... 1 1.1 Applications ..........................1 1.1.1 Intended Use ........................1 1.1.2 Summary of the Instrument ....................1 1.1.3 Principles of Operation ...................... 3 1.2 Warning Markings Inscriptions d’avertissement ..............4 1.2.1 Safety Symbols Le Symboles de Sûreté ................4 1.2.2 Safety Terms Terminologie de Sûreté...
Page 4
4.4.10 4-Parameter and 5-Parameter Logistic ................. 50 4.4.11 % Absorbance Mode ..................... 51 5. ChroMate ® Manager Software ..............52 5.1 Opening ChroMate Manager ....................52 ® 5.1.1 Toolbar Icons ........................52 5.2 Management Menu ........................ 53 5.2.1 Test Count ........................54 5.2.2 New Assay Calibration Data ...................
Page 5
6.2 Errors and Flags ........................80 6.3 COM Port Settings ......................... 82 6.3.1 PC Communication (COM) Port Setting ................82 6.3.2 Communication (COM) Port Setting ................85 7. Contact Information ..................86 8. Appendix A – Laboratory Information System (LIS) ........87 9.
1.1 Applications 1.1.1 Intended Use ® is a compact, PC-controlled, multi-purpose photometer system designed to ChroMate read and calculate the results of assays, which are read in 96 well microtiter plates or strip trays. General purpose software is included. ChroMate ®...
Using bichromatic differential absorbance values corrects for optical imperfections in the plastic wells and removes the effects of meniscus and turbidity. The photo-detectors generate electronic signals which are amplified and interpreted by the instrument circuitry. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
BIOHAZARD BIOHAZARDS are biological agents that can cause disease in humans. Lab workers handling potentially infectious materials must use universal precautions to reduce the risk of exposure to these agents. CAUTION! L’ATTENTION! ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
WARNING: Treat all components during use and disposal as you would any biohazardous material. AVERTISSEMENT : Utiliser et disposer des matériaux de la même manière que vous utilisé et disposer des matières infectieuses. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Do Not Operate Without WARNING: Do not operate this instrument with covers and panels Protective Covers removed. Avoid Exposed Circuitry WARNING: Do not touch exposed connections and components when power is present. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Page 13
If needed, a mild all-purpose or nonabrasive cleaner may be used. Use as a disinfectant a 10% solution of chlorine bleach or 70% • isopropyl alcohol. Take special care not to spill liquid inside the instrument • Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Avertissement! Lors du fonctionnement, si on renverse des matériaux, coupez immédiatement le courant. Placez le commutateur électrique a AU LOIN(0). Traitez le matériel comme biohazardous, utilisant approprie nettoient et des méthodes de disposition. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
PC Connection: ........ USB Port Microprocessor: ........ eZ80 Speed: Reads absorbance of 96 wells in approximately 20 seconds Plate Transport: ........ Stepper motor Certifications: ........CE, ETL listed, NRTL listed (USA/CAN), Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 16
Technical Specifications (Continued) • Software: PC interface with ChroMate ® Manager software included (PC not included) Features: .......... PC interface software (included) with complete user prompting, error messages, full plate reporting, automatic interpretation options, duplicate options, controls locator, curve plotting and editing.
If power fluctuations or loss are frequent, an Uninterruptible Power Supply (UPS) is recommended. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 18
Avertissement! Pour la protection continue contre le risque du feu, remplacez FUSIBLE le fusible seulement par une du type spécifique et des estimations courantes. Démontez l’équipement de l’alimentation d’énergie avant de remplacer le fusible. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Figure 3.3-2 Well A1 location D. Lamp access door E. Universal Serial Bus (USB) connection F. External Protective Grounding Terminal G. Power switch I=ON O=OFF H. Power Cord Connection Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 20
Figure 3.3-3 Back of Instrument ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Turn on the computer and insert the installation CD or USB drive. The installer should start automatically. If not, or if the software is on a USB flash drive, locate the software by searching from the Windows start button. Double click to open the ChroMate ®...
Page 22
There are 2 important steps to take when installing and accessing ChroMate Manager software on Windows 8 or 10 computer systems. You must follow the steps below for all of the software ® features to be properly installed. When installing the Manager software, right click on the setup.exe file and select “Run As Administrator”...
Driver installation - if installation of the software does not complete, minimize the installation windows in order to click OK on this screen: The PC COM port setting does not require an adjustment before activating ChroMate ® Manager. However, instructions on how to view and adjust the setting are provided in Section 6.3.1 PC Communication (COM) Port Setting.
Follow the instructions under the Instrument Setup & Software Installation Section 3.3 to ensure that the ChroMate ® Manager software is installed and the USB cable is attached. Push the power switch to turn ON the computer and the instrument. Then double click the ChroMate Manager ® icon to start the software.
10. Will a cut off value be used to label negative samples or define an equivocal zone? If so, results less than what cut off should be labeled as negative? Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
To use the plate map, a test must be programmed and stored. To read raw absorbance without programming a test, use the "Read Plate Absorbance" mode located under the Tests option on the main toolbar ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
If the instrument firmware version is lower than AE:26, pressing the Mix button shows this message "Instrument firmware version does not support this feature. Please update the firmware!". Check with your instrument supplier. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
4.2 Running an Assay The ChroMate Manager Tabs are located at the bottom of the screen: ® Tab Title: Description: Item # Plate Layout Plate Layout is the default window. Refer to Section 4.2.1. Assay Refer to Section 4.2.2 Calibration Add Samples Refer to Section 4.2.3...
Read Plate button to run. Figure 4.2.1-1 Plate Layout Tab Figure 4.2.1-1 displays the current status of the instrument. Also shown is the currently loaded plate. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 30
(a total of 96 wells). Important note: assays run, or samples run against stored curves, must all have the same filter configuration when run in the same plate. Remove Removes well assignments. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Figure 4.2.1.1-1 Manual Plate Assignment 4.2.2 Assay Calibration Tab CAUTION! CAUTION: Errors or alterations to calibration curves may introduce error to test results. ChroMate features ® are available for user convenience but may not safely apply to all test applications.
Page 32
Assay Calibration Tab (Continued) Figure 4.2.2-2 Assay Calibration Tab NOTE: When ChroMate is finished reading, select New Assay Calibration Data from ® Management on the toolbar, or by pressing the F8 key on your keyboard. This window shows whether the new curves and controls were accepted.
Page 33
(as shown above), click on the Accept Selected button to accept. The stored curve is displayed in the window. The stored curve will then be used for sample calculations. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
“Choose an assay” box to display the curve. If the curve was valid, it will be automatically accepted and shown in the lower right box as below. Figure 4.2.2.2-1 Graph of a curve ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
(and generate a new curve) by adding the new calibrators and blanks. You should answer “yes” at a minimum when a new reagent lot is being used. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 36
Select the assay from the drop down list (Figure 4.2.2.3-3) and then click on the graph icon (1) and the stored calibration curve will display on the right hand side of the screen (2). Figure 4.2.2.3-3 Example Standard Curve Example ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Page 37
Moving the mouse pointer over the plotted curve will display the curve fit parameters. 4.2.2.3-4 View Curve fit parameters Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
If there are sample results, they will be updated on the Test Results tab. You can now choose to accept or delete any of the sample results by clicking on the buttons. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
NOTE: Decreasing the standard curve by a percentage will result in increasing the sample concentration values. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Standard 5 is read individually. Figure 4.2.2.6-1 Adjusting a curve with one or more Standards By selecting this standard and all the stored standards, the new adjusted curve can be activated. Figure 4.2.2.6-2 Adjusted Curve ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
4.2.2.7 View assay parameters The parameters of an assay can be viewed by hovering the mouse cursor over the drop down arrow at the end of the window that displays the assay. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Click on one or more of the numeric/sample ID’s on the left Tests to highlight, choose a test to run with the chosen IDs. Click on the test and press ‘Add Test’. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
NOTE: Do not use zero for dilution factor. The Dilution Factor is not stored with the plate data. If a Plate is stored and reloaded, the dilution factor must be set again. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 44
Test Results and Report tabs will use MDF=x.xxx under column Notes to indicate what dilution factor was used. If the Add to Plate button has been selected, the screen will be directed to Plate Layout. Figure 4.2.3.1-2 Dilution Factor can be reported under Notes ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Click ‘Accept’ to accept the results of the selected ID. buttons Press ‘Delete’ to discard the selected ID. NOTE: Results must be accepted to allow them to be viewed in the Report Tab. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Check this box to select all of the results. Only the selected results will appear in printouts. Sorting Select the Sorting Option from the drop down menu to Option change the way the results are displayed. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Page 47
(*.xls), or XML file (*.xml). Save for future reference. Remove Selecting the Remove button will remove all selected items from the results display. Retest Adds the selected items to the Sample Tab to be re-tested. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Select the Offset Absorbance field and enter the user-determined offset absorbance value. Press the Save button to save the settings. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
4.4 Assay Modes software contains several pre-programmed general-purpose calculations selected to ChroMate ® facilitate data handling for enzyme immunoassays and other similar tests. Access the Assay Modes via ChroMate ® Manager Tests Menu (refer to Section 5.2.2) •Absorbance •Factor •Standard •Cut Off...
The instrument first accepts a calibrator read singly or in multiple numbers and then calculates concentrations based on a single- point standard curve passing through the point (0,0). A blank is required to determine the (0,0) point. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
• Use this equation if only a negative control is used to determine the COV. COV = Y * mean (Pc) +F • Use this equation if only a positive control is used to determine the COV. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 52
(positive <= cut off, negative > cut off). The instrument calculates the mean of the positives and the mean of the negatives. User programmable QC Criteria is available on all controls. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
8.“Y=Log(1000*ABS), X=Log(Conc)” -- the base-10 log 0.0 concentration Calibrator is of the absorbance is plotted against the base-10 used, it will be ignored when log of the concentration the regression is calculated.) Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
The resulting calibrator curve is a smooth curve connecting the calibrator points, which may be entered in ascending or descending order of absorbance. A constraining algorithm is applied to prevent curve overshoot. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Fourth order is defined by: Y = A + B * X + C * X^2 + D * X ^3 + E * X ^4 Graphing and editing options are available. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
(concave upwards to concave downwards). = Asymptote at infinite concentration: in an ELISA assay where you have a standard curve, this can be thought of as the response value for infinite standard concentration. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
The instrument accepts a number of calibrators read singly or in multiple numbers. Blanking, graphing, and editing options are available. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Power on the computer. Power on the instrument and wait for the initialization to complete. Double-click on the ChroMate icon to open the software. ® NOTE: It is not necessary to turn the instrument off when restarting the software.
5.2 Management Menu Figure 5.2-1 ChroMate ® Manager Menu Bar Figure 5.2-2 Management Menu Option Description Connect Establishes or re-establishes communication between the software on the PC to the instrument. Communication Opens the interactive Communication Window used for certain Window diagnostics.
5.2.2 New Assay Calibration Data When the instrument is finished reading, the New Assay Calibration Data can be viewed by ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Blanks are menus used. Blank in wells data will display in the Plate Data table when the Read Plate button is selected. Figure 5.3.1-1 Read Plate Absorbance Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
30 seconds, the instrument will not mix between reads. Read intervals need to be greater than 30 seconds and the ‘Mix between reads’ box checked before the instrument will mix between reads. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Raw Data worksheet, (row/column A1) followed by the second run (row/column A11) and so on. 12 x 8 Way checkbox allows user to define export format. • Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Use the ‘Properties’ button to make any adjustments to the Concentration or Minimum number of the standards entered. Figure 5.3.2.2-1 Assign Standard Properties If reading standards are read in duplicate, triplicate, etc., enter the number in the “Minimum number to use” field. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
‘Issue Warning’ or ‘Invalidate Tests’ as action to take when result is out of range, and to set the Valid Time, in days/hours. If no entry is made for Day(s) or Hour(s), expiration will be set at the default of (7) days. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Set up the Out of Range Action and the Valid Time the Control(s).If no entry is made for Day(s) or Hour(s), expiration will be set at the default of (7) days. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Click the Positive/Negative button and specify the ‘Positive greater than or equal to’ value, and ‘Negative less than’ value. Units may be typed in or drop down menu may be used. Figure 5.3.2.8-3 Positive/Negative ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
The ‘Print’ button sends data to a printer. • Save Assay The ‘Save’ button saves assay settings. 5.3.3 Edit Assay Edit Assay is accessed via the Tests Menu selection on the ChroMate Manager Menu ® Bar: Figure 5.3.3-1 Select Stored Assay Select the Assay from the drop down menu.
Identify the Substance to be registered by clicking on the substance title from the dropdown menu. Click “Register” Enter the lot number, expiration date and description in the Note field (optional). Click OK to Save. Figure 5.4-1 Lot Registration window ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
NOTE: Lot numbers are entered in the tab labeled Lot # Registration. Edit by selecting the Select QC Points button or by un-checking the checkbox for specific QC points. Select the Save QC Points button to save the QC points displayed. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Modify a Sample ID record by clicking on the field to edit and typing. Delete a Sample ID by highlighting the record to delete and press the Delete • key on the keyboard. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Highlight the items to be deleted then press the Delete key on the keyboard. A dialog box will open prompting the user to confirm the deletion of the record. Press Yes to delete, press No to cancel the deletion process. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Select the ‘Import’ button at the bottom of the Sample Database window. Select ‘Open’ to locate the text file to be imported. Select the data fields to be imported. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Enter criteria into Search Database area, press OK. All records containing the specified criteria will display. All records in the Patient Database can be displayed by pressing OK and leaving all search fields blank. Search result for “Smoker” displayed Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
COM Port setting to Refer to match the COM Port Section 6.2 setting of the COM Port computer Settings Refer to Section 4.2.3.1 Refer to Section 5.7.1.1.1 International Sample Dilution Factor Calibration Factor (ICF) ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Click the OK button once all the selections have been made. NOTE: The Sample ID field and the Keyword field must be used. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Print logo and page header must be checked to have to include these features on reports. Logo and font size can be edited. A printed space can be added for notes and signatures. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Report Type window, allows less paper to be Instead of one test or patient per page, multiple results will print on the allowable paper space. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Tab and preview button. Preview sample reports via the Report Tab and preview button. Modify the example Custom Report and save it to a new file name. Figure 5.7.2.1-1 Custom Report Sample ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
A list of the available language choices will display. After selecting a language, it will be necessary to close the Manager and restart the software to initialize the selected language. To open the ChroMate Manager software, right click on the ChroMate icon and select “Run as Administrator”.
Acceptable voltage readings are within in the “greater than 2.0” and “less than 10.0” range. These values are for AFV (amplified filter voltages.). Blue range is < 7V Green range is 7 to 9V Red range is > 9V Figure 5.8.1-2 Voltage Meter ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Figure 5.8.2-1 Password Required Figure 5.8.2-2 Firmware Update 5.8.3 Calibration Lock/Unlock The calibration settings for the ChroMate ® are set by the factory. The Calibration Lock/Unlock feature is locked at the factory as well. This feature is for emergency use only by authorized personnel in case the device would need to be recalibrated.
5.9 Help The Help ‘About’ option provides a dialog window that displays the version of ChroMate ® Manager software and the version of firmware that is currently installed. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Customer Service”, it is very helpful for you to provide a file along with your email. Follow these instructions to create the “Root.zip” file: Click on "My Computer" and go to C:\ChroMate. • Right click on the file labeled "Root", select "Send To" and then "Compressed •...
Error "flags" appear as codes in the lower right hand corner of the screen. A flag is temporary and users may sometimes miss them. You can navigate to C:\ChroMate\Root\Logs on your PC and open the “MgrErr.log”. This log file will contain a history of any error messages that have appeared as well as an error code number.
Page 87
"Run as Administrator". If the error message returns, it may be the operating system security or the location of the setup file. Be sure to copy the ChroMate installation folder to the hard drive (C:) of the computer.
6.3 COM Port Settings Generally the communication (COM) port setting of the PC and the ChroMate will not require ® adjustment. However, the following steps are included just in case the user needs to view the setting or make an adjustment to the setting.
Page 89
Click on the Ports (COM & LPT) folder. You should see a version of Silicon driver . If Silicon Labs is NOT shown, recheck the connection between the instrument and the PC and be sure the instrument’s power switch is set to I (ON). Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 90
14 from the drop down list of options.in the Manager software. Be sure to select OK to exit at the bottom of the window. ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Click to highlight the setting and then click on the OK button to save the setting. NOTE: The COM port setting will remain at this setting until it is changed. Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
1935 SW Martin Hwy. Palm City FL 34990 USA Important: When contacting us, please have the Model and Serial Number of the ChroMate ® in question. Have a description of the problem with as much detail as possible. Save any relevant jobs or logs and send or e-mail us the information.
Results from the Requested Assay preceding. Indicates the End of the file. Append this information to the preceding record; may be used on any record. Figure 8.1-1 Record Type Definitions Microplate Reader Operator’s Manual Doc. 4300 Rev. Q ChroMate ®...
Page 94
NOTE: A solid bar (|) indicates the field is for future use. The bar delimiter is necessary for correct formatting within the record. File Examples: Figure 8.1-2 Import File Example (.LIS file extension) (.LIS file extension) Example Figure 8.1-3 Export ChroMate ® Microplate Reader Operator’s Manual Rev. Q...
Using the computer’s Control Panel, uninstall any previous version of ChroMate ® Manager. Locate and double click on the ChroMate software folder. Right click on the setup.exe file and “Run as Administrator”. Follow the on screen instructions to install the software.