7.0 Trouble-shooting
Problem
Check
The chair
Mains cable is connected?
cannot be
Main switch is on?
moved
Is the green control lamp illuminated?
Is the green LED on the board "PWR"
switched on and is a beep audible
during switch on?
Voltage at the plug CN9
(24–28 V AC on the board
near the small transformer)
Problem
Check
Foot control
Observe operating instructions.
pedal does
not work
properly
Despite perfect operation, the foot control pedal
does not always work properly.
Problem
Check
Chair only
Does the relay on the board work properly?
moves
upwards and
Measure the voltage at the motor (24 – 28 V AC).
then it stops
Problem
Check
Chair can
Check the connector on the board.
be moved
but position
storage is
Check the button for position memory.
not possible
Problem
Check
Trendelen-
Does the relay on the board work properly?
burg motor
Measure the voltage at the motor (24 – 28 V AC).
does not
work
Problem
Check
Deviation
Correct adjustment of requested position?
from the
chair
position
stored
Problem
Check
Motor
Loud cracking and sometimes the
makes
motor does not work.
noises
16

Trouble shooting

Connect the mains cable to the mains supply.
Switch on the mains switch.
Check the fuses, 2 pieces 3,15 A,
at the power input.
Check the connector on the board. Fuse on the board
T 315 mA and T 12,5 A.
Check the fuse F 12,5 A between transformer and board.
Fuse between transformer and board F 12,5 A or
exchange transformer.
Trouble shooting
Press the control pedal straight and not skewed.
In order not to twist the pedal, observe the delay regarding
the position memory.
Exchange the board in the foot pedal and check perfect pass
of the cable, the way the cover fits perfectly.
Trouble shooting
If not, exchange the board.
If voltage is connected and the motor does not work,
exchange the motor.
Trouble shooting
Rectify the fault (exchange the connector or the board).
If required, exchange the switch.
Trouble shooting
If not, exchange the board.
If voltage is connected and the motor does not work,
exchange the motor.
Trouble shooting
Bring the chair into "0" position, then adjust the requested
position and store it, as described in the operating
instructions.
Trouble shooting
Probably the gear mechanism is defective, therefore
exchange the motor.