7.0 Maintenance
7.2

Sending in the device

1.
Remove and properly dispose of consumables.
2.
Clean and disinfect the products and accessories according to the operating instructions.
3.
Place used accessories with the device.
4.
Fill in the form QD 434 „Delivery complaint / return shipment" and the respective Decontamination certificate.
• This form is enclosed to each delivery and can be found at www.atmosmed.com.
5.
The device must be well padded and packed in suitable packaging.
6.
Place the form QD 434 „Delivery complaint / return shipment" and the respective decontamination certificate in an
envelope.
7.
Affix the envelope to the outside of the package.
8.
Send the product to ATMOS or to your dealer.

8.0 Trouble-shooting

The ATMOS
C 361 was subjected to a thorough quality control before shipment. If there is, nevertheless, some malfunction,
®
you possibly might solve this problem yourselve if you observe the following instructions:
Problem
 Unit does not start
 Insufficient performance
or no suction
Possible causes
– Loose power plug
– No power voltage
– Defective fuse
– Leakages within the hose system or in
the secretion canister lid
– Filter is clogged
(vacuumgauge indicates a vacuum)
– Secretion or blood has been sucked in
and valve plates of the pump are
contaminated sealed?
Remedy
– Check connection to supply socket
– Check inbuilding fuse
– Replace fuse
– Check secretion canister lid and hose
system, replace sealing ring on
secretion canister lid, if necessary
– Replace filter, check filling level in
secretion canister; evacuate canister,
if necessary
– Unit has to be returned for repair