Summary of Contents for Atmos Chair 41 Gyne

  • Page 1 MedizinTechnik English ATMOS Chair 41 Gyne 503.0550.B 2019-07 Index: 20...
  • Page 2: Table Of Contents

    Consumables ............... 18 9.0 Technical specifications ..........19 10.0 Disposal ..............20 11.0 Notes on EMC ............21-23 ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Str. 16 79853 Lenzkirch Deutschland/Germany Phone: + 49 7653 689-0 Fax: + 49 7653 689-392 (Sales Germany)
  • Page 3 General survey This survey also shows options*, which are not part of the basic equipment! MAGNETKISSEN MAGNETIC PILLOW COUSSIN MAGNÉTIQUE GRIFF HANDLE POIGNEÉ BEINHALTER LEG SUPPORT ECARTEUR RÜCKENLEHNE BACKREST DOSSIER SITZFLÄCHE SEAT SIÉGE KOLPOSKOPHALTER COLPOSCOPE ARM BRAS POUR COLPOSCOPE SEKRETSCHALE BASIN CUVETTE FUßSTÜTZE...
  • Page 4: Introduction

    General information The ATMOS Chair 41 Gyne does not electromagnetically or otherwise affect other equipment nor is the chair itself affected by these phenomena, caused by other equipment, as it is manufactured in compliance with the Directive, and meets the requirements of the harmonised standard...
  • Page 5: Intended Use

    - Max. lifting capacity: 200 kg - Prior to cleaning, the mains switch must be switched off. - The ATMOS Chair 41 Gyne may be operated only in rooms used for medical purposes, but not in areas subject to explosion hazards and in oxygen rich environments.
  • Page 6: Main Characteristics

    Introduction This chapter describes the features of the gynaecological chair ATMOS Chair 41 Gyne. The main equipment components and relative accessories are also illustrated. Warning: This equipment has been conceived for gynaecology purposes only. Neither the chair nor its components ...
  • Page 7: Type Label

    Introduction Type label The type label is situated on the rear part of the chair and indicates the model, the serial number, the REF and other information (see Fig. 1.1). Fig. 1.1 Technical information Fig. 1.2...
  • Page 8: Scope Of Supply

    For removing the packing, all screws must be loosened and the strap must be removed.  The following components must be supplied inside the packing of the bottom part: • 1 bottom part of the ATMOS Chair 41 Gyne • 1 operating instructions • 1 bottle of detergent •...
  • Page 9: Setting Up And Starting Up

    3.0 Setting up and starting up Loosen the two screws that secure the chair to the wooden pallet to free it. A second person is required to help to move the chair from the pallet. Use the transport handles to move the chair. Use the lateral handles at the seat to position the upper part onto the lower part.
  • Page 10: Setting Up And Starting Up

    3.0 Setting up and starting up Once the two parts of the chair have been assembled and placed in the desired position, securing it, if necessary to the floor, exploiting the holes in the base (see Fig. 3.2), proceed as follows: 1.
  • Page 11: Operation

    Thanks to the integrated control electronics, the chair has many functions, which are easy to use. All the chair adjustments and functions are co-ordinated, processed and stored by an integrated microprocessor. The ATMOS Chair 41 Gyne can easily be controlled when using the foot pedal, with it the gynaecologist has his hands free for the patient.
  • Page 12: Programming

    4.0 Operation Warning: When moving the chair make sure that the patient has her hands at the chair and that third persons and things are  in adequate distance from the chair. Programming Proceed as follows to program (keys “I”,”II” and “III” of the right-hand foot pedal): 1.
  • Page 13: Cleaning And Care

    5.0 Cleaning and care In case the chair is left inoperative for a longer period, it is advisable to cover it with a sheet, to switch it off and disconnect it from the mains supply. Cleaning For a longer service life, cleaning with a wet cloth is recommended at regular intervals. Upholstery For removing stains on the upholstery, only the supplied detergent may be used.
  • Page 14: Recommended Surface Disinfectants

    5.0 Cleaning and care Fig 5.1 Recommended surface disinfectants The surfaces of the Chair 41 Gyne can be cleaned with disinfectants with the following active ingredients: • quaternary ammoniumchlorides 5.3.1 Recommended disinfectant for cushion Disinfectant Contents (in 100 g) Manufacturer Green &...
  • Page 15: Maintenance And Servicing

    To carry out these measures the person must have the necessary test devices and original spare parts. ATMOS recommends: work should be carried out by an authorized ATMOS service partner. This ensures that repairs and testing are carried out professionally, original spare parts are used and warranty claims remain unaffected.
  • Page 16: Trouble Shooting

    7.0 Trouble-shooting Problem Check Trouble shooting The chair Mains cable is connected? Connect the mains cable to the mains supply. cannot be Main switch is on? Switch on the mains switch. moved Is the green control lamp illuminated? Check the fuses, 2 pieces 3,15 A, at the power input.
  • Page 17: Accessories, Further Options, Consumables

    Accessories, further options and consumables Accessories   Headrest magnetic pillow 503.0556.0 Colour: Pastel blue 503.0556.1 Colour: Saffron yellow Colour: Salmon 503.0556.2 503.0556.3 Colour: Leaf green 503.0556.4 Colour: Light grey 503.0556.5 Colour: Silk grey 503.0556.6 Colour: White 503.0556.7 Colour: Sapphire blue ...
  • Page 18: Consumables

    Colour: White  503.0570.7 Colour: Sapphire blue 503.0553.0  Colposcope holder mountable left or right, for colposcope für ATMOS i View Kaps support with tube diameter from 30 - 38 mm; Consumables for the ATMOS Chair 41 Gyne Paper role 503.0557.0...
  • Page 19: Technical Specifications

    Technical specifications Nominal voltage 230 V ± 10% Nominal frequency 50-60 Hz Fuses 2 × T 3.15 A 250 V Control voltage Motor voltage 24 V Conditions +10...+35°C Transport and Storage 20...80 % atmospheric pressure from 500...1060 hPa max. seat height (mm) min.
  • Page 20: Disposal

    10.0 Disposal ● The ATMOS Chair 41 Gyne is not comprised of any hazardous materials. ● The materials of the housing can be recycled completely. ● Prior to disposal, device and accessories must be decontaminated. ● The materials are to be separated carefully.
  • Page 21: Notes On Emc

    Guidelines and Manufacturer´s Declaration - Emissions The ATMOS Chair 41 Gyne is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS Chair 41 Gyne should ensure that it is used in such an environment.
  • Page 22 The ATMOS Chair 41 Gyne is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS Chair 41 Gyne should ensure that it is used in such an environment.
  • Page 23 If the measured field strength at the location where the ATMOS Chair 41 Gyne is used exceeds the above compliance level, the ATMOS Chair 41 Gyne is to be observed to verify the intended use. If abnormal performance characteristics are noted, additional measures might be necessary, e.g. a changed arrangement or another location for the device.
  • Page 24 ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Str. 16 79853 Lenzkirch / Germany Phone: +49 7653 689-0 atmos@atmosmed.de www.atmosmed.com...