6.0 Maintenance and servicing

Maintenance, repairs and period tests may only be carried out by persons who have the appropriate technical knowledge
and are familiar with the product. To carry out these measures the person must have the necessary test devices and
original spare parts.
ATMOS recommends: work should be carried out by an authorized ATMOS service partner. This ensures that repairs and
testing are carried out professionally, original spare parts are used and warranty claims remain unaffected.
At least every 24 months a repeat test of the electrical safety should be performed according to IEC 62353.
ATMOS recommends an inspection according to the manufacturer's specifications.
6.1

Exchange of upholstery

Seat upholstery
This upholstery is fixed with nylon snap fasteners and can therefore easily be removed in pulling it off. If required, adjust
the lower and the upper snap fastener for a perfect fit of the upholstery.
Back upholstery
This upholstery is fixed with magnets and nylon snap fasteners. For removing the upholstery, loosen the fixation screws
of the upholstery support, which is fixed at the hinges near the seat.