Alarm Conditions

A.2.5
Priority
Medium
Low
Note:
During multiple-alarm conditions, alarms are presented as follows:
If two medium-priority alarms occur simultaneously, the alarm status message toggles between the
two.
If a medium-priority and low-priority alarm occur simultaneously, only the medium-priority alarm is
displayed and only the medium-priority tone is sounded.
A medium-priority physiological alarm overrides a low-priority alarm.
Operator's Manual
Table A-5.  Alarm Conditions
Rate
Sounds every 8 seconds
Yellow
while the alarm condi-
Steady message, slow
tion persists
flashing label
Sounds every 16 seconds
Steady yellow
while the alarm condi-
tion persists
Color
Type
Physiological
Technical
Technical
Specifications
Messages
High rSO
2
Low rSO
2
Sensor not connected
Excessive light
Poor signal quality
Preamp not connected
Replace sensor
Interference detected
Sensor unusable
Battery low
COM port unavailable
Printer unavailable
USB storage full
USB storage error
Auto baseline set
A-3