Specifications
A.7
Environmental
The following environmental conditions shall be observed:
Table A-12. Environmental Conditions for Storage or Transport
Temperature
-40 °F to +158 °F
(-40° C to +70° C)
Table A-13. Environmental Conditions for Operation
Temperature
+41 °F to10 4 °F (+5°C to 40°C)
Under extreme conditions of use that are beyond the recommendations above but within the limits of
a supply voltage of –20%, compared to the nominal temperature or the combination of a temperature
of 113 °F (45 °C) and humidity of 75% HR, the ventilator should not malfunction nor endanger the
user. However, operating the device for prolonged periods or repeatedly under such extreme
conditions could result in premature aging of components and more frequent maintenance.
A.8
Pneumatic
Table A-14. Airway Resistances
Inspiratory
1.0 cmH
O at 30 lpm flow
2
3.7 cmH
O at 60 lpm flow
2
Table A-15. Patient Circuit Resistances
Adult Double Limb
≤
2 cmH
O at 60 lpm flow
2
a. Includes exhalation valve
b. Values obtained from the manufacturer's (DAR) directions for use.
Table A-16. Air Inlet Resistance (Filter)
0.057 cmH
O (5.6 pascals)
2
a. All testing conducted on 3-inch diameter
test piece
Table A-17. Oxygen Inlet Specifications
Maximum pressure
7.25 psi (50 kPa)
A-12
Humidity-40 °F to 158 °F
(-40° C to +70° C)
10% to 95% RH
Humidity
10% to 95%
RH
Expiratory
0.5 cmH
O at 30 lpm
2
1.1 cmH
O at 60 lpm
2
a
Pediatric Double Limb
b
≤
2 cmH
O at 30 lpm flow
2
a
at 24 lpm flow
Maximum flow
15 lpm
Atmospheric
Altitude
pressure
7.2 psi to 15.4 psi
-500ft to 13,000 ft
(50 kPa to 106 kPa)
(-152 m to 3964 m)
Atmospheric
Altitude
pressure
8.7 psi to 16.0 psi
-500ft to 13,000 ft
(60 kPa to 110 kPa)
(-152 m to 3964 m)
Puritan Bennett 540™ Ventilator User's Manual