III CLEANING AND REPROCESSING
Follow the recommendations of the "Reprocessing Instructions" (FB-358/NA) manual delivered with your product
regarding procedure for cleaning and sterilizing the components, and the present-day regulations on reprocessing in effect in
your country.
Reusable Products:
Description
AIRFLOW
MAX handpiece
CLIP+CLEAN
®
PERIOFLOW
handpiece
Easy Clean
®
Scan
for
support
3.1- Water lines cleaning & disinfection
If you have the AIRFLOW
STATION+, bottle has to be placed onto the cart to run the lines. To disinfect the
®
unit place the CLEANER bottle on the bottom only.
Keeping the device's water lines clean and disinfected is mandatory to prevent patient infection.
WATER SUPPLY
BY HOSE EG-110
A regular cleaning and maintenance protocol should be adopted to clean and protect dental unit
NIGHT
waterlines. EMS recommends using EPA-registered dental unit waterline cleaners, VistaCleanTM
CLEANER
Irrigant Solution Concentrate and VistaTab™ Dental Unit Waterline Cleaner by Hu-Friedy or
Monarch Lines Cleaner by Air Techniques.
Follow the instructions for use of the product.
The instructions for use should be followed to ensure the appropriate water quality to help protect
patients, staff and equipment.
Handpiece should be removed prior to using VistaCleanTM /VistaTab™ or Monarch Lines
Cleaner.
The water supply hose and related device connection will not be cleaned by this procedure.
Initial start-up treatment and routine treatment
Follow the instructions for use of the product.
For daily water line disinfection: Add drops
16
of VistaCleanTM Irrigant Solution Concentrate in the water bottle fully filled (800ml)
and use it with patients.
Dental unit waterline cleaner and AIRFLOW
ONE
®
1- Waterlines cleaning with VistaTab™ Dental Unit Waterline Cleaner or Monarch Lines Cleaner.
Follow the instructions for use of the product.
1
2
CLEANER
Place the CLEANER bottle onto
Set water to 10.
the device with VistaTab™ or
Turn the device ON.
Monarch Lines Cleaner
Set water regulator to 10 to
ensure the flow of the cleaning
agent.
16 Follow disinfection products manufacturer instructions. Please refer to your public health guidelines.
16
3
4
Press the pedal once,
release, and wait 1 minute.
Hold cord over a sink.
Repeat if necessary until the
solution is visible.
Contamination prevention:
DO NOT make any contact
between the sink and the
cords.
2- Flushing
1
2
H
O
2
Place the water bottle fully
Set water to 10.
filled.
Turn the device ON.
To reduce the risk of
Set water regulator to 10 to
ingestion of the cleaning
ensure optimal rinsing.
agent by the patient, always
use a fully filled 800ml
water bottle.
DO NOT leave water in your water bottles overnight. Allow the bottles to air with the lids off overnight.
Always empty out and wash the water bottle used for rinsing before any new use. EMS recommends a weekly use of a bottle
cleaning agent.
After every patient
Overall cleaning and disinfection
1
Use a cleaning and disinfecting wipe (less than 35% alcohol) compatible EPA-
registered intermediate level surface disinfectant or that complies with the standards
applicable in the country (e.g. MICROKLEEN or Opti-Cide³ wipes).
Follow carefully the instructions provided by the wipes manufacturer.
DO NOT use CaviWipes™ or Advantaclear wipes. They damage EMS products
3
4
Press the pedal until the water
Hold cord over a sink.
runs clear.
Contamination prevention:
Cleaning can be paused and
DO NOT make any contact
reset by pressing the pedal
between the sink and the
again.
cords.
Remove the water bottle and
rinse well prior to filling with
water for the first patient of the
day.
2
Reprocess
handpieces
and
instruments. Please refer to FB-
358/NA "reprocessing instructions"
delivered with your kit.
Risk of contamination. Always
disinfect the bottom and top areas of
device air connections.
Please remember to dry your
AIRFLOW
cord connection and ensure
®
it is cooled and dried prior to each use.
Excess moisture can cause blockages.
17