4.4.2. Protective class II equipments equipped with mains plug

Substitute leakage current between mains part and exposed conductive part of the
tested equipment should be tested.
We advice to use aluminium foil about 200 mm x 100 mm for testing of insulation
resistance on equipments, which have no exposed conductive part. In such case
coat checked part with foil and connect it to one pole of test voltage.
Take into consideration all safety warnings!
Do not touch exposed conductive parts of tested object (including foil) while
measurement is in progress - risk of electric shock! Symbol „ " is displayed
on LCD if the measurement is in progress!
Step 1
Press the I
LEAK
Either result of last measurement or „---" will be displayed.
Step 2
Plug in the mains cable of equipment under test to Test socket 1 (Fig. 8).
Switch on the main switch of equipment under test.
Connect test lead P 2011 with test tip P 3011 to
equipment under test with test lead.
Step 4
Press the START key and keep it pressed, until displayed result is stabilized, then
release the key. Symbol „ " indicates, that measurement is running.
Copyright © 2004 ILLKO, s.r.o.
SONDE
TEMP.
PRÜFDOSE / TEST SOCKET
R
/
I
ISO LEAK
5
I
PE INV.
ESC 2051
I
6
DIFF.
PE
DEL
I
7
DIFF. INV.
LOAD
DEL
ESC
P
8
VA
START
SAVE
CAL
ENTER
9
MENU
key and release it to select substitute leakage current function.
SONDE
TEMP.
R
R
/
I
PE
ISO LEAK
R
0
5
I
PE
PE INV.
200mA
I
U/R
1
6
DIFF.
PE
10A
DEL
I
2
7
R
DIFF. INV.
ISO
LOAD
P
3
I
8
VA
LEAK
SAVE
4
I
9
MENU
PE
Fig. 8
terminal. Touch conductive part of
25
ESC 2051
PRÜFDOSE / TEST SOCKET
ESC 2051
DEL
ESC
START
CAL
ENTER

Advertisement

Table of Contents
loading

Table of Contents