5.2. Description of MENU items

„Last values"
This item allows latest measured values examining and identification number and
protection class of tested equipment examining and editing. The procedure of identification
number and protection class entering/editing is described in chapter 5.3.1.
Browsing with ▲ and ▼ keys through submenu „Last values" displays last measured
values, which are saved in the buffer. „Val" symbol is displayed, too.
„Memory"
This item allows browsing through records saved in instrument's main memory. When you
browse with ▲ and ▼ keys through records in memory, identification number of particular
tested equipment is displayed. After you press the ENTER key and release it, you can
browse through measured values of selected particular equipment. „Mem" symbol is
displayed, too. Selected record can be transferred to the buffer and then checked or edited.
Detailed description of memory operation is described in chapter 5.3.2.
„Set-up"
This item allows changing of some system settings. Once it is set it remains in instrument's
memory even if it is disconnected from mains.
„I external"
This item allows switching the method of measurement of leakage currents, power and
current. You can switch between „Yes" and „No" with the ENTER key. If „ Yes" is
selected then „Device current" message is displayed (clamp current measurement), if you
press the P
key. Enabling of „I external" is indicated by displaying „X" symbol, too.