Step 1 - preparation for measurement
•
Select the I
SUB
and indicators (I
•
Connect tested appliance:
One-phase appliance with movable power supply cord – plug in its mains plug
-
to test socket 1.
Three-phase appliance with movable power supply cord – connect PE pin of its
-
plug to the instrument's PE terminal by means of optional test lead P2012 +
crocodile clip P4012. Then connect remaining current-carrying conductors by
means of test lead P2011 + test tip P3011 to the instrument's R
terminal (placed on rear panel).
Hard-wired appliance – firstly disconnect appliance from mains! Connect its
-
point of PE connection to the instrument's PE terminal by means of optional
test lead P2012 + crocodile clip P4012. Then connect remaining current-
carrying conductors by means of test lead P2011 + test tip P3011 to the
instrument's R
Step 2 – measurement
•
Switch on tested appliance's mains switch.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
•
control details). Measurement is indicated by the (START) indicator.
•
After finishing of measurement (the (START) indicator turns off) the last measured
value of substitute leakage current is displayed.
4.4.2. Protective class II appliances – substitute touch leakage current
1
Connect tested appliance; its mains switch must be
2
switched on
Press the test tip connected to the I
3
against exposed conductive part of tested appliance
4
Copyright © 2006–2019 ILLKO, s.r.o.
ST ART
function: pres the I
SUB
) and (mA) light up.
SUB
+ I
terminal (placed on rear panel).
ISO
SUB
I
SUB
ST ART
ST ART
Fig. 7 – Substitute touch leakage current measurement
Fig. 6 – Substitute leakage current measurement
key and release it. Symbol „- -" is displayed
M
K
A
A
V
m
W
VA
cos
terminal
D
18
REVEXplus, REVEXplus USB
Measurement
+ I
ISO
SUB
Function selection
Appliance connecting
Measurement