Press the ENTER key and release it.
Step 2
Use instrument's numerical keyboard to enter identification number. You can use the
DEL key to erase incorrectly entered characters.
Use the ENTER key to store entered identification number or use the ESC key to
cancel entering of identification number.
Editing of tested equipment's protective class
Protective class of tested equipment is set automatically. If R
or 10 A), protective class I is set, otherwise class II i set.
In order to change protective class proceed as follows:
Step 1
After you have finished all measurements of particular equipment, press the MENU
key and release it.
Use ▲ and ▼ keys to select submenu „ Last values" and press ENTER.
Use ▲ and ▼ keys to select item „Class" and press ENTER.
Step 2
Press the ENTER key and release it repetitively to select appropriate protective class I,
II or III.
Press the ESC key and release it (repeatedly) until you abandon (sub)menus.
Buffer erasing
Measurement result must be displayed. Press the DEL key and keep it pressed. Press
the ENTER key. Release both keys.

5.3.2. Main memory

Transferring buffer content to main memory
User can use two ways to transfer buffer content to main memory. It is possible to use either
"quick storing" or "standard storing".
All records are placed ascending in main memory in accordance with identification number, it
means from the lowest to the highest identification number.
If you are storing record with the identification number, under which was former stored other
record, the older record will be overwritten with the new one!
Copyright © 2004 ILLKO, s.r.o.
Ident. number
----------
Val
was measured (either 200 mA
PE
53
ESC 2051