Troubleshooting
Description
█
PCA
10:00 - 70%
Air or Up
Occlusion
Check for ai r / Occlusion
Press
key to mute
█
PCA
10:00 - 70%
Down
Occlusion
Check Set / Access
Press
key to mute
█
PCA
10:00 - 70%
Pump Paused
Too Long
Press
key
█
PCA
10:00 - 70%
Door Open
Close door
█
PCA
10:00 - 10%
Low Battery
Connect to Charger
BodyGuard 595™Color Vision Operator Manual
Result
Possible Cause
Infusion stops
Air is present in administration
■
and an alarm is
Set.
activated
The roller or clip on the
administration set is closed
upstream of the pump.
The line was not loaded correctly.
Infusion stops
The roller or clip on the
■
and an alarm is
administration set is closed
activated
downstream of the pump.
Administration set is kinked.
Cannula is blocked.
Administration set loaded
incorrectly.
NOTE: operator may change
pressure setting.
An alarm is
Two minutes has elapsed without
■
activated
a button press during
programming.
Infusion stops
The door of the pump was not
■
and an alarm is
closed prior to operation
activated
The door has accidentally opened
during operation.
Pump will only
30 Minut es of battery life are
■
run for another
remaining.
30 minut es if it is
not connected to
mains.
5 – Alarm Conditions and Troubleshooting
Required Action
Disconnect line from patient, press
key. Pump stopped screen is
displayed. Press & hold (3sec) on the
key, then select Prime option.
Follow the steps in "Priming a set"
section.
Re-load the administration line.
Open clamp/clip.
Straighten the set.
Change the Cannula.
Re-load the administration line.
Refer to "Down Occlusion Alarm"
section.
Press
to resume.
Close the door of the pump.
Send the pump for service.
Place the pump in charger and
connect to mains.
63