To Unlock Keypad
Press and hold
beep is heard. The beep indicates that the locking
program is turned OFF.
NOTE: to prevent unauthorized individuals accessing the main menu a technique
is required to return to the menu code screen when the pump is in STOP mode.
To return to the menu during infusion, press
the STOP state.
Press and hold
NOTE: when keypad lock is on, entering the STOP state & holding down
key will not return you to the main menu. Press
infusion, press and hold
hold
key until you are prompted for the level one access code to return to
the main menu.
Dedicated Administration Sets
The BodyGuard 595 Color Vision™ Pain Manager should be operated only with
TM
MicroSet
dedicated Infusion sets, deferent color for deferent program: PCA (Blue),
Epidural (Yellow) and Regional (Green).
Warning: do not operate the BodyGuard 595 Color Vision™ Pain Manager
with any administration set, other than the MicroSet
infusion sets are equipped with a free flow protection valve, to prevent a
gravity free flow hazard. Using a different set may essentially change the
delivering rate and expose the patient to a free flow rate, which can cause
serious injury or death.
Warning: using the pump with air in line detector off, may cause an embolism
resulting death or paralysis. Switching the air detector off is not recommended
and should be used only when absolutely necessary. Please contact your
technician in order to enable the air sensor if disabled with no need.
BodyGuard 595™Color Vision Operator Manual
until the entire graph is clear and a
key until the level one access code prompt appears.