2 – Symbols, Warnings and Cautions
Watch your fingers and nails when opening the pump door.
When operati ng the pump on PCA program with a rate of 0.0 ml/h there is a
hazard of blood clot forming. Connect saline infusion in parallel to avoid this
problem.
When operating the pump on PCA program, the program will not enter KVO,
even if the limit volume is acceded. The program will complete the bolus and
than enter KVO.
In a clinical setting, bolus cables can be damaged by crushing (under foot or in
the bed frames) or cutting (trapped in the bed-frame or cut by scissors when
removing dressings, etc.)
BodyGuard 595™Color Vision Operator Manual
18