CME BodyGuard 595 Color Vision Operator's Manual page 39

Pain manager infusion system
Table of Contents

Advertisement

25. Set bolus dose volume and press
26. Set a maximum bolus dose allowed for the
protocol
.
27. Set bolus interval and press
28. Set a maximum interval time between boluses in
for a protocol. Press
29. In case of patient bolus combined with auto
bolus: patient will periodically receive an
automatic bolus until the hour limit or VTBI have
been reached. Press
BodyGuard 595™Color Vision Operator Manual
.
.
to continue.
key.
PCA
10:00
A: MORPHINE
Auto-Bolus Dose
3
mL
Enter and Press
Key
PCA
10:00
A: MORPHINE
Maximum Bolus Dose
5
mL
Enter and Press
Key
PCA
10:00
A: MORPHINE
Auto-Bolus interval
2
Minutes
Enter and Press
Key
PCA
10:00
A: MORPHINE
Maximum Auto-Bolus interval
1
Minutes
Enter and Press
Key
PCA
10:00
A: MORPHINE
Warning !
Patient will
Periodically receive
an automatic bolus.
Press
to Confirm
4 – Operation
-
70%
-
70%
-
70%
-
70%
-
70%
39